The SHELLY neonatal incubator provides a protected and comfortable environment for newborn patients, from standard to intensive care. It combines thermal comfort, acoustic comfort, dual microprocessor control, vital function monitoring and an integrated monitoring display.
Key Features
- Low noise level in the patient space: 38 dB.
- Automatic air curtain activation when the incubator door is opened to reduce heat loss.
- Click-in double-walled cover for greater stability and an additional acoustic barrier.
- Two-stage light and acoustic alarm signalling with on-screen cause and advice.
- Dual microprocessor control of all functions.
- Integrated Masimo Rainbow SET® monitoring platform for SpO2, perfusion index, pulse rate, PVi and SpHb.
- Integrated high-precision scales with weight trends and increments.
- Adjustable bed, extensible to both sides and rotatable by 45° at the patient’s head, with an X-ray cassette drawer.
- 9-inch colour LED display with positioning and rotation adjustment.
- Servo-controlled air humidification with low-water indication and autoclave sterilisation capability.
- IRIS ports, 5 port holes, 10 hose inputs and two entrance doors tiltable by 180°.
- Euro rail, lower shelf and optional storage drawers.
Technical Specifications
- Type designation: SI-610-1
- Power voltage: 110/120/230 V, 50–60 Hz
- Power input: 550 VA fixed height; 600 VA adjustable height
- Dimensions: Height 1450 mm; adjustable undercarriage 1320–1520 mm at top cover and 1575–1775 mm at IV pole; length 1150 mm; width 610 mm
- Weight: 96 kg; 106 kg with height-adjustable undercarriage
- Noise level: maximum 38 dB inside the canopy without O2 dosing
- Electric bed positioning: Trendelenburg and anti-Trendelenburg, 12°
- Air temperature mode: 20.0–37.0 °C and 37.1–39.0 °C, in 0.1 °C increments
- Body temperature mode: 34.0–37.0 °C and 37.1–39.0 °C, in 0.1 °C increments
- Temperature measurement accuracy: ±0.1 °C ±1 digit
- Temperature control accuracy: ±0.3 °C from the adjusted value
- O2 concentration adjustment: 22–75%, in 1% increments; display range 10–100%
- O2 measurement accuracy: ±3% ±1 digit; regulation accuracy ±1.5% O2 from the adjusted value
- Relative humidity adjustment: 40–95%, in 1% increments; display range 10–99%
- Attainable relative humidity: 95%; measurement and regulation accuracy ±5%
- Water tank: 1.5 l demineralized water
- Weighing range: maximum 6 kg; resolution 1 g
- Scales accuracy: ±2 g ±1 digit (0–1 kg), ±5 g ±1 digit (1–3 kg), ±10 g ±1 digit (3–6 kg)
- SpO2 displayed range: 0–100%; measurement accuracy ±3% ±1 digit within 70–100% SpO2
- Pulse rate: displayed range 25–240 bpm; accuracy ±3 bpm
- PI displayed range: 0–20%; accuracy ±10% of PI
- PVi displayed range: 0–100%
- SpHb displayed range: 0–25 g/dL; measurement accuracy ±1 g/dL
Optional Accessories
Vital function monitoring with Masimo SpO2, dual temperature probes, oxygen regulation, integrated weighing scale, double Plexiglas wall, hose holder, examination light, IRIS hand port, infusion hook, additional shelves, closed storage boxes, backup power, pedal options and antistatic wheels.